%0 Journal Article %T Development and validation of ultra-performance liquid chromatographic method with tandem mass spectrometry for the determination of lenalidomide in rabbit and human plasma %A Muzaffar Iqbal %A Tanveer A Wani %A Nasr Y Khalil %A Ibrahim A Darwish %J Chemistry Central Journal %D 2013 %I BioMed Central %R 10.1186/1752-153x-7-7 %X A simple, highly sensitive UPLC-MS/MS method was developed and validated for the determination of LND in rabbit and human plasma. After a simple protein precipitation using methanol, LND and carbamazepine (IS) were separated on Acquity UPLC BEHTM C18 column (50 x 2.1 mm, i.d. 1.7 mum, Waters, USA) using a mobile phase consisted of acetonitrile:water:formic acid (65:35:0.1%, v/v/v) pumped at a flow rate of 0.2 mL/min. LND and IS were eluted at 0.71 and 1.92 min, respectively. The mass spectrometric determination was carried out using an electrospray interface operated in the positive mode with multiple reaction monitoring (MRM) mode. The precursor to product ion transitions of m/z 260.1 > 149.0 and m/z 237.0 > 179.0 were used to quantify LND and IS, respectively. The method was linear in the concentration range of 0.23--1000 ng/mL with a limit of quantitation of 0.23 ng/mL. All the validation parameters were in the ranges acceptable by the guidelines of analytical method validation.The proposed UPLC-MS/MS method is simple, rapid and highly sensitive, and hence it could be reliable for pharmacokinetic and toxicokinetic study in both animals and humans. %U http://journal.chemistrycentral.com/content/7/1/7/abstract