%0 Journal Article %T Interference between pacemakers/implantable cardioverter defibrillators and video capsule endoscopy %A Dirk Bandorski %A Johannes Gehron %A Reinhard H£¿ltgen %J World Journal of Gastrointestinal Endoscopy %D 2013 %I Baishideng Publishing Group Co. Limited %R 10.4253/wjge.v5.i4.201 %X Our Letter to the Editor, related to the article ¡°Small bowel capsule endoscopy in patients with cardiac pacemakers and implantable cardioverter defibrillators: Outcome analysis using telemetry¡± by Cuschieri et al, comments on some small errors, that slipped into the authors discussions. The given informations concerning the pacemaker- and implantable cardioverter defibrillators modes were inaccurate and differ between the text and the table. Moreover, as 8 of 20 patient¡¯s pacemakers were programmed to VOO or DOO (¡°interference mode¡±) and one patient was not monitored by telemetry during capsule endoscopy, 9 of 20 patients (45%) lack the informations of possible interference between capsule endoscopy their implanted device. Another objection refers to the interpretation of an electrocardiogram (figure 1, trace B) presented: in contrast to the author¡¯s opinion the marked spike should be interpreted as an artefact and not as ¡±undersensing of a fibrillatory wave¡±. Finally, three comments to cited reviews were not complete respectively not quoted correctly. %K Capsule endoscopy %K Small bowel capsule endoscopy %K Interference %K Cardiac pacemaker %K Implantable cardioverter defibrillator %K Telemetry %U http://www.wjgnet.com/1948-5190/full/v5/i4/201.htm