%0 Journal Article %T Dosimetric Characteristics of 6 MV Medical Linac at BAEC %A Sudeb Kumar Roy %A Pretam Kumar Das %A Rajada Khatun %A Md. Ashikur Rahman %A Shirin Akter %A Tushar Kumar %A Mohammad Monjur Ahasan %J International Journal of Medical Physics,Clinical Engineering and Radiation Oncology %P 38-46 %@ 2168-5444 %D 2021 %I Scientific Research Publishing %R 10.4236/ijmpcero.2021.101004 %X Dosimetric characteristic is one of the essential parameters of a medical linear accelerator (LINAC), which must be obtained before clinical use. The dosimetric characteristics for 6 MV photon beam were measured and compared with the corresponding published data. The study was done using a Varian linear accelerator (Model Clinac-iX) at the Institute of Nuclear Medical Physics (INMP), AERE, Savar, Dhaka, under the Bangladesh Atomic Energy Commission (BAEC). The data is taken for 10 field sizes (2 กม 2, 3 กม 3, 5 กม 5, 7 กม 7, 10 กม 10, 15 กม 15, 20 กม 20, 25 กม 25, 30 กม 30 and 40 กม 40 cm2) at same conditions. The measured Percent Depth Dose (PDD) curves were obtained for 6 MV photon beams with the field as mentioned above and compared with the calculated PDD curves. The measured depth dose (Dmax) for reference field size (FS) 10 กม 10 cm2 is 15.99 mm, and the PDD at 10 cm depth (D10) is 66.87% for 6 MV photon energies that are found to be compatible with the published report BJR supplement 25. The measured PDD curves for photon energies show a good agreement with the standard PDD curves. The photon beam dosimetry data found in the current study are compatible and all the tolerances are within the clinically acceptable tolerance limit. %K Linac Machines %K 3D Water Phantom %K Ionization Chamber %K Electrometer %U http://www.scirp.org/journal/PaperInformation.aspx?PaperID=107136