Accuracy and Reliability of Uterine Contraction Identification Using Abdominal Surface Electrodes

DOI: 10.4137/CMWH.S10444

Full-Text   Cite this paper   Add to My Lib

Abstract:

Objective: To compare the accuracy and reliability of uterine contraction identification from maternal abdominal electrohysterogram and tocodynamometer with an intrauterine pressure transducer. Methods: Seventy-four term parturients had uterine contractions monitored simultaneously with electrohysterography, tocodynamometry, and intrauterine pressure measurement. Results: Electrohysterography was more reliable than tocodynamometry when compared to the intrauterine method (97.1 versus 60.9 positive percent agreement; P < 0.001). The root mean square error was lower for electrohysterography than tocodynamometry in the first stage (0.88 versus 1.22 contractions/10 minutes; P < 0.001), and equivalent to tocodynamometry in the second. The positive predictive values for tocodynamometry and electrohysterography (84.1% versus 78.7%) were not significantly different, nor were the false positive rates (21.3% versus 15.9%; P = 0.052). The sensitivity of electrohysterography was superior to that of tocodynamometry (86.0 versus 73.6%; P < 0.001). Conclusion: The electrohysterographic technique was more reliable and similar in accuracy to tocodynamometry in detecting intrapartum uterine contractions.