Comparative dosimetry of GammaMed Plus high-dose rate 192 Ir brachytherapy source

Keywords: 0.1-cc chamber , AAPM TG-43 , dosimetry , EGSnrc simulation code , high dose rate

Full-Text   Cite this paper   Add to My Lib

Abstract:

The comparative dosimetry of GammaMed (GM) Plus high-dose rate brachytherapy source was performed by an experiment using 0.1-cc thimble ionization chamber and simulation-based study using EGSnrc code. In-water dose measurements were performed with 0.1-cc chamber to derive the radial dose function (r = 0.8 to 20.0 cm) and anisotropy function (r = 5.0 cm with polar angle from 10° to 170°). The nonuniformity correction factor for 0.1-cc chamber was applied for in-water measurements at shorter distances from the source. The EGSnrc code was used to derive the dose rate constant (L), radial dose function g L (r) and anisotropy function F(r, q) of GM Plus source. The dosimetric data derived using EGSnrc code in our study were in very good agreement relative to published data for GM Plus source. The radial dose function up to 12 cm derived from measured dose using 0.1-cc chamber was in agreement within ±3% of data derived by the simulation study.