核磁共振谱自动基线校正新方法

DOI: 10.11938/cjmr20140407, PP. 523-534

Keywords: 核磁共振(NMR),自动基线校正,迭代方法,数据处理

Full-Text   Cite this paper   Add to My Lib

Abstract:

该文提出一种新的核磁共振谱自动基线校正方法,该方法基于迭代移动平均值算法,改进了峰区域的基线识别,然后使用迭代方法进行基线建模.为了改进峰区域的基线建模结果,该方法结合了两种基线校正方法的优点,在基线识别过程选择了一系列位于峰区域的“基线定位点”,然后和除了峰区域以外的基线点一起参与基线建模.该基线校正方法适用于NMR谱,即使在基线严重非线性失真的情况下也能得到理想的基线校正结果

References

[1]  Hoch J C, Stern A S. NMR Data Processing[M]. New York: Wiley-Liss, 1996.
[2]  Tang C G. An analysis of baseline distoration and offset in NMR spectra[J]. J Magn Reson, 1994, 109(2): 232-240.
[3]  Marion D, Bax A. Baseline distoration in real fourier transform NMR spectra[J]. J Magn Reson, 1988, 79(2): 352-356.
[4]  Wang Chao(王超), Huang Ying-ying(黄颖颖), Yang Guang(杨光). Automatic phase correction for NMR spectra -connecting line of equally spaced points(clesp) (核磁共振谱图自动相位校正新方法——等密度点连线法)[J]. Chinese J Magn Reson(波谱学杂志) , 2004, 21(4): 445-457.
[5]  Huang Ying-ying(黄颖颖), Li Peng(李鹏), Liu Xiao-zheng(刘小征), et al. A synthetical algorithm for automatic phase correction of NMR spectra based on neural network (NNAPC)(基于神经网络的NMR 谱图自动相位校正综合算法)[J]. Chinese J Magn Reson(波谱学杂志) , 2007, 24(1): 43-51.
[6]  Yuan Qing-quan(袁清泉), Guo Jun-wang(郭俊旺), Cong Jian-bo(丛建波), et al. An EPR data acquisition and control system for in vivo tooth dosimetry(在体剂量检测专用EPR 数据采集与控制系统)[J]. Chinese J Magn Reson(波谱学杂志) , 2013, 30(3): 354-360.
[7]  Moskau D. Application of real time digital filters in NMR spectroscopy[J]. Concept Magn Reson, 2002, 15(2): 164-176.
[8]  Wider G. Elimination of baseline artifacts in NMR-spectra by oversampling[J]. J Magn Reson, 1990, 89(2): 406-409.
[9]  Eilers P H C. A perfect smoother[J]. Anal Chem, 2003, 75(14): 3631-3636.
[10]  Cobas J C, Bernstein M A, Martin-Pastor M, et al. A new general-purpose fully automatic baseline-correction procedure for 1D and 2D NMR data[J]. J Magn Reson, 2006, 183(1): 145-151.
[11]  Bao Q J, Feng J W, Chen F, et al. A new automatic baseline correction method based on iterative method[J]. J Magn Reson,2012, 218: 35-43.
[12]  Zhao J H, Lui H, McLean D I, et al. Automated autofluorescence background subtraction algorithm for biomedical Raman spectroscopy[J]. Appl Spectrosc, 2007, 61(11): 1 225-1 232.
[13]  Prakash B D, Wei Y C. A fully automated iterative moving averaging (AIMA) technique for baseline correction[J]. Analyst, 2011, 136(15): 3 130-3 135.
[14]  Johnsen L G, Skov T, Houlberg U, et al. An automated method for baseline correction, peak finding and peak grouping in chromatographic data[J]. Analyst, 2013, 138(12): 3 502-3 511.
[15]  Shao X G, Ma C X. A general approach to derivative calculation using wavelet transform[J]. Chemometr Intell Lab, 2003, 69(1-2): 157-165.
[16]  Golotvin S, Williams A. Improved baseline recognition and modeling of FT NMR spectra[J]. J Magn Reson, 2000, 146(1): 122-125.
[17]  Dietrich W, Rudel C H, Neumann M. Fast and precise automatic base-line correction of one-dimensional and 2-dimensional NMR-spectra[J]. J Magn Reson, 1991, 91(1): 1-11.