全部 标题 作者
关键词 摘要

OALib Journal期刊
ISSN: 2333-9721
费用:99美元

查看量下载量

相关文章

更多...
-  2017 

A Device for Ribbon Collection for Array Tomography with Scanning Electron Microscopy

DOI: 10.1267/ahc.17013

Keywords: array tomography, three-dimensional observation, serial ultrathin section, post-embedding labeling, scanning electron microscopy

Full-Text   Cite this paper   Add to My Lib

Abstract:

“Array tomography” is a method used to observe the fine structure of cells and tissues in a three-dimensional view. In this method, serial ultrathin sections in the ribbon state (ribbons) are mounted on a solid substrate and observed by scanning electron microscopy (SEM). The method may also be used in conjunction with post-embedding immunocytochemistry. However, it is difficult to mount many serial ribbons on a substrate manually. We developed an inexpensive laboratory-made device that mounts ribbons by pulling a nylon fishing line and lifting the substrate up from the water in a knife boat. Using this device, we succeeded in mounting several ribbons consisting a mean of 205.6 (SD: 37.7) serial ultrathin sections on 1.25 (SD: 0.06) × 1.25 (SD: 0.06)-cm silicon substrates. Furthermore, it was confirmed that our method is suitable for ribbons derived from water-soluble resin blocks. We were also able to stain the specimens by post-embedding immunocytochemistry. Thus, our method is useful in mounting numerus sections on a substrate for array tomography with SEM

Full-Text

Contact Us

[email protected]

QQ:3279437679

WhatsApp +8615387084133