Comparison of FASTMAP and B0 Field Map Shimming at 4T: Magnetic Field Mapping Using a Gradient-Echo Pulse Sequence

DOI: 10.4236/jemaa.2020.128010, PP. 115-130

Keywords: MRI, Shimming, 4T, Higher-Order Shims

Full-Text   Cite this paper   Add to My Lib

Abstract:

Local susceptibility variations result in B0 field inhomogeneities, causing distortions and signal losses in MR imaging. Susceptibility variations become stronger with increasing B0 magnetic field strength. Active shimming is used to generate corrective magnetic fields, which can be used to improve B0 field homogeneity. FASTMAP is an effective shimming technique for computing optimal coil currents, which uses data from six projection directions (or columns): this technique is routinely used for shimming cubic volumes of interest (VOIs). In this paper, we propose several improvements to FASTMAP at 4T. For each shim coil, using a modified 3D gradient-echo pulse sequence, we compute B0 inhomogeneity maps and project them onto eight 1st and 2nd order spherical harmonic functions. This process is repeated for shim currents between -15,000 to 15,000 with increments of 5000 Digital to Analog Converter (DAC) units, and is used to compute the gradient between spherical harmonic coefficients and DAC values for all 8 shim coils—along with the R2 values of linear fits. A method is proposed (based on R2 values) to further refine optimal shim currents in respective coils. We present an analysis that is numerically robust and completely flexible in the selection of the VOIs for shimming. Performance analyses, phantom results, and in vivo results of a human brain are presented, comparing our methods with the FASTMAP method.

References

[1]  Gruetter, R. and Tkáč, I. (2000) Field Mapping without Reference Scan Using Asymmetric Echo-Planar Techniques. Magnetic Resonance in Medicine, 43, 319-323.
https://doi.org/10.1002/(SICI)1522-2594(200002)43:2%3C319::AID-MRM22%3E3.0.CO;2-1
[2]  Sumanaweera, T.S., Glove, G.H., Binford, T.O. and Adler, J.R. (1993) MR Susceptibility Misregistration Correction. IEEE Transactions on Medical Imaging, 12, 251-259.
https://doi.org/10.1109/42.232253
[3]  Holz, D., Jensen, D., Proksa, R., Tochtrop, M. and Vollmann, W. (1988) Automatic Shimming for Localized Spectroscopy. Medical Physics, 15, 898-903.
https://doi.org/10.1118/1.596173
[4]  Doddrell, D.M., Galloway, G.J., Brewton, I.M. and Brooks, W.M. (1988) Nodal Inhomogeneity Mapping by Localized Excitation—The “NIMBLE” Shimming Technique for High-Resolution in Vivo NMR Spectroscopy. Magnetic Resonance in Medicine, 7, 352-357.
https://doi.org/10.1002/mrm.1910070313
[5]  Hoult, D. (1987) Shimming on Spatially Localized Signals. Journal of Magnetic Resonance, 73, 174-177.
https://doi.org/10.1016/0022-2364(87)90238-1
[6]  Haacke, E.M., et al. (1999) Magnetic Resonance Imaging: Physical Principles and Sequence Design. Wiley-Liss, New York.
[7]  Wilson, J.L., Jenkinson, M. and Jezzard, P. (2002) Optimization of Static Field Homogeneity in Human Brain Using Diamagnetic Passive Shims. Magnetic Resonance in Medicine, 48, 906-914.
https://doi.org/10.1002/mrm.10298
[8]  Juchem, C., et al. (2006) Combined Passive and Active Shimming for in Vivo MR Spectroscopy at High Magnetic Fields. Journal of Magnetic Resonance, 183, 278-289.
https://doi.org/10.1016/j.jmr.2006.09.002
[9]  Cusack, R., et al. (2005) An Evaluation of the Use of Passive Shimming to Improve Frontal Sensitivity in fMRI. Neurolmage, 24, 82-91.
https://doi.org/10.1016/j.neuroimage.2004.08.029
[10]  Koch, K.M., et al. (2006) Sample-Specific Diamagnetic and Paramagnetic Passive Shimming. Journal of Magnetic Resonance, 182, 66-74.
https://doi.org/10.1016/j.jmr.2006.06.013
[11]  Klassen, L.M. and Menon, R.S. (2004) Robust Automated Shimming Technique Using Arbitrary Mapping Acquisition Parameters (RASTAMAP). Magnetic Resonance in Medicine, 51, 881-887.
https://doi.org/10.1002/mrm.20094
[12]  Jayatilake, M., Storrs, J. and Lee, J.-H. (2008) Comparison of FASTMAP and B0 Field Map Shimming for 4T MR. University of Cincinnati, Ohio.
[13]  Gruetter, R. (1993) Automatic, Localized in Vivo Adjustment of All First-And Second-Order shim coils. Magnetic Resonance in Medicine, 29, 804-811.
https://doi.org/10.1002/mrm.1910290613
[14]  Gruetter, R. and Boesch, C. (1992) Fast, Noniterative Shimming of Spatially Localized Signals. In Vivo Analysis of the Magnetic Field along Axes. Journal of Magnetic Resonance, 96, 323-334.
https://doi.org/10.1016/0022-2364(92)90085-L
[15]  Miyasaka, N., Takahashi, K. and Hetherington, H.P. (2006) Fully automated Shim Mapping Method for Spectroscopic Imaging of the Mouse Brain at 9.4 T. Magnetic Resonance in Medicine, 55, 198-202. https://doi.org/10.1002/mrm.20731
[16]  Hsu, J.J. and Glover, G.H. (2005) Mitigation of Susceptibility-Induced Signal Loss in Neuroimaging Using Localized Shim Coils. Magnetic Resonance in Medicine, 53, 243-248.
https://doi.org/10.1002/mrm.20365
[17]  Mackenzie, I.S., Robinson, E.M., Wells, A.N. and Wood, B. (1987) A Simple Field Map for Shimming. Magnetic Resonance in Medicine, 5, 262-268.
https://doi.org/10.1002/mrm.1910050307
[18]  Prammer, M.G., et al. (1988) A New Approach to Automatic Shimming. Journal of Magnetic Resonance (1969), 77, 40-52.
https://doi.org/10.1016/0022-2364(88)90030-3
[19]  Wen, H. and Jaffer, F.A. (1995) An in Vivo Automated Shimming Method Taking into Account Shim Current Constraints. Magnetic Resonance in Medicine, 34, 898-904.
https://doi.org/10.1002/mrm.1910340616
[20]  Shen, J., Rothman, D.L., Hetherington, H.P. and Pan, J.W. (1999) Linear Projection Method for Automatic Slice Shimming. Magnetic Resonance in Medicine, 42, 1082-1088.
https://doi.org/10.1002/(SICI)1522-2594(199912)42:6%3C1082::AID-MRM12%3E3.0.CO;2-G
[21]  Chen, Z., et al. (2004) Measurement and Automatic Correction of High-Order B0 Inhomogeneity in the Rat Brain at 11.7 Tesla. Magnetic Resonance Imaging, 22, 835-842.
https://doi.org/10.1016/j.mri.2004.01.062
[22]  Roméo, F. and Hoult, D.I. (1984) Magnet Field Profiling: Analysis and Correcting Coil Design. Magnetic Resonance in Medicine, 1, 44-65.
https://doi.org/10.1002/mrm.1910010107
[23]  Shen, J., Rycyna, R.E. and Rothman, D.L. (1997) Improvements on an in Vivo Automatic Shimming Method (FASTERMAP). Magnetic resonance in Medicine, 38, 834-839.
https://doi.org/10.1002/mrm.1910380521
[24]  Glover, G. and Schneider, E. (1991) Three-Point Dixon Technique for True Water/Fat Decomposition with B0 Inhomogeneity Correction. Magnetic Resonance in Medicine, 18, 371-383.
https://doi.org/10.1002/mrm.1910180211
[25]  Reynaud, O., et al. (2015) Fast Low-Specific Absorption Rate B0-Mapping along Projections at High Field Using Two-Dimensional Radiofrequency Pulses. Magnetic Resonance in Medicine, 73, 901-908.
https://doi.org/10.1002/mrm.25217
[26]  Zhang, Y., Li, S. and Shen, J. (2009) Automatic High-Order Shimming Using Parallel Columns Mapping (PACMAP). Magnetic Resonance in Medicine, 62, 1073-1079.
https://doi.org/10.1002/mrm.22077
[27]  Jayatilake, M., et al. (2011) Construction and Optimization of Local 3rd Order Passive Shim System for Human Brain Imaging at 4T MRI. Proceedings of the International Society for Magnetic Resonance in Medicine, 19, 3785.
[28]  Jackson, J.D. (1999) Classical Electrodynamics. 3rd Edition, Willey, Hoboken.
[29]  Arfken, G.B. and Weber, H.J. (1999) Mathematical Methods for Physicists. 4th Edition, Academic Press, Cambridge.
[30]  Kim, D.H., Adalsteinsson, E., Glover, G.H. and Spielman, D.M. (2002) Regularized Higher-Order in Vivo Shimming. Magnetic Resonance in Medicine, 48, 715-722.
https://doi.org/10.1002/mrm.10267
[31]  Webb, P. and Macovski, A. (1991) Rapid, Fully Automatic, Arbitrary-Volume in Vivo Shimming. Magnetic Resonance in Medicine, 20, 113-122.
https://doi.org/10.1002/mrm.1910200112
[32]  Kanayamay, S., Kuhara, S. and Satoh, K. (1996) In Vivo Rapid Magnetic Field Measurement and Shimming Using Single Scan Differential Phase Mapping. Magnetic Resonance in Medicine, 36, 637-642.
https://doi.org/10.1002/mrm.1910360421

Full-Text

comments powered by Disqus

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133