Purpose: The purpose of this study is to provide technical information on commissioning the TrueBeam STx 6 MV flattening-filter free (FFF) beam in the RayStation treatment planning system (TPS) for stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) treatments. Methods: For beam modeling, percent depth dose curves, profiles and output factors for jaw-collimated fields and stereotactic cones as well as X-jaws transmission were measured. For multi-leaf collimator (MLC) modeling, MLC model parameters such as offset, gain, curvature, leaf tip width, tongue and groove and transmission were determined and output factors for MLC-collimated fields were measured. Absolute dose calibration was also performed. For beam model and MLC model validation, the American Association of Physicists in Medicine Task Group-119 plans, clinical SRS and SBRT plans and end-to-end testing were performed. Results: Beam characteristics of the 6 MV FFF beam agreed well with those in the literature. Validation results showed that our beam model and MLC model were acceptable for SRS and SBRT treatments. Conclusions: The technical information and dosimetric data provided in this study will be a useful reference for other clinics/institutions which will commission the same machine energy in the RayStation TPS.
References
[1]
Fogliata, A., Fleckenstein, J., Schneider, F., Pachoud, M., Ghandour, S., Krauss, H., et al. (2016) Flattening Filter Free Beams from TrueBeam and Versa HD Units: Evaluation of the Parameters for Quality Assurance. Medical Physics, 43, 205.
https://doi.org/10.1118/1.4938060
[2]
Chang, Z., Wu, Q., Adamson, J., Ren, L., Bowsher, J., Yan, H., et al. (2012) Commissioning and Dosimetric Characteristics of TrueBeam System: Composite Data of Three Truebeam Machines. Medical Physics, 39, 6981-7018.
https://doi.org/10.1118/1.4762682
[3]
Narayanasamy, G., Saenz, D., Cruz, W., Ha, C.S., Papanikolaou, N. and Stathakis, S. (2016) Commissioning an Elekta Versa HD Linear Accelerator. Journal of Applied Clinical Medical Physics, 17, 179-191. https://doi.org/10.1120/jacmp.v17i1.5799
[4]
Thomas, E.M., Popple, R.A., Prendergast, B.M., Clark, G.M., Dobelbower, M.C. and Fiveash, J.B. (2013) Effects of Flattening Filter-Free and Volumetric-Modulated Arc Therapy Delivery on Treatment Efficiency. Journal of Applied Clinical Medical Physics, 14, 155-166. https://doi.org/10.1120/jacmp.v14i6.4328
[5]
Glide-Hurst, C., Bellon, M., Foster, R., Altunbas, C., Speiser, M., Altman, M., et al. (2013) Commissioning of the Varian TrueBeam Linear Accelerator: A Multi-Institutional Study. Medical Physics, 40, Article ID: 031719.
https://doi.org/10.1118/1.4790563
[6]
Beyer, G.P. (2013) Commissioning Measurements for Photon Beam Data on Three TrueBeam Linear Accelerators, and Comparison with Trilogy and Clinac 2100 Linear Accelerators. Journal of Applied Clinical Medical Physics, 14, 273-288.
https://doi.org/10.1120/jacmp.v14i1.4077
[7]
Chen, S., Yi, B.Y., Yang, X., Xu, H., Prado, K.L. and D’Souza WD. (2015) Optimizing the MLC Model Parameters for IMRT in the RayStation Treatment Planning System. Journal of Applied Clinical Medical Physics, 16, 322-332.
https://doi.org/10.1120/jacmp.v16i5.5548
[8]
Savini, A., Bartolucci, F., Fidanza, C., Rosica, F. and Orlandi, G. (2016) Optimisation and Assessment of the MLC Model in the Raystation Treatment Planning System. Radiotherapy and Oncology, 119, S380-S381.
https://doi.org/10.1016/S0167-8140(16)32056-4
[9]
Saez, J., Hernandez, V., Goossens, J., Kerf, G.D. and Verellen, D. (2020) A Novel Procedure for Determining the Optimal MLC Configuration Parameters in Treatment Planning Systems Based on Measurements with a Farmer Chamber. Physics in Medicine & Biology, 65, Article ID: 155006.
https://doi.org/10.1088/1361-6560/ab8cd5
[10]
Das, I.J., Cheng, C.W., Watts, R.J., Ahnesjö, A., Gibbons, J., Li, X.A., et al. (2009) Accelerator Beam Data Commissioning Equipment and Procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM. Medical Physics, 35, 4186-4215. https://doi.org/10.1118/1.2969070
[11]
Li, J., Lu, B., Yan, G., Huang, Y. and Liu, C. (2012) SU-D-BRCD-02: A Novel Cylindrical 3D Water Scanner for Beam Data Collection: I. Validation of the Automatic-Setup. Medical Physics, 39, 3613-3613. https://doi.org/10.1118/1.4734668
[12]
Pönisch, F., Titt, U., Vassiliev, O.N., Kry, S.F. and Mohan, R. (2006) Properties of Unflattened Photon Beams Shaped by a Multileaf Collimator. Medical Physics, 33, 1738-1746. https://doi.org/10.1118/1.2201149
[13]
IAEA and AAPM. (2017) Dosimetry of Small Static Fields Used in External Beam Radiotherapy. Technical Reports Series No. 483. International Atomic Energy Agency, Viena.
https://www.iaea.org/publications/11075/dosimetry-of-small-static-fields-used-in-external-beam-radiotherapy.
[14]
Tanny, S., Sperling, N. and Parsai, E.I. (2015) Correction Factor Measurements for Multiple Detectors Used in Small Field Dosimetry on the Varian Edge Radiosurgery System. Medical Physics, 42, 5370-5376. https://doi.org/10.1118/1.4928602
[15]
Smilowitz, J.B., Das, I.J., Feygelman, V., Fraass, B.A., Kry, S.F., Marshall, I.R., Mihailidis, D.N., et al. (2015) AAPM Medical Physics Practice Guideline 5.a.: Commissioning and QA of Treatment Planning Dose Calculations—Megavoltage Photon and Electron Beams. Journal of Applied Clinical Medical Physics, 16, 14-34.
https://doi.org/10.1120/jacmp.v16i5.5768
[16]
Almond, P.R., Biggs, P.J., Coursey, B.M., Hanson, W.F., Huq, M.S., Nath, R., et al. (1999) AAPM’s TG-51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams. Medical Physics, 26, 1847-1870.
https://doi.org/10.1118/1.598691
[17]
McEwen, M., DeWerd, L., Ibbott, G., Followill, D., Rogers, D.W.O., Seltzer, S., et al. (2014) Addendum to the AAPM’s TG-51 Protocol for Clinical Reference Dosimetry of High-Energy Photon Beams. Medical Physics, 41, Article ID: 041501.
https://doi.org/10.1118/1.4866223
[18]
Ezzell, G.A., Burmeister, J.W., Dogan, N., LoSasso, T.J., Mechalakos, J.G., Mihailidis, D., et al. (2009) IMRT Commissioning: Multiple Institution Planning and Dosimetry Comparisons, a Report from AAPM Task Group 119. Medical Physics, 36, 5359-5373. https://doi.org/10.1118/1.3238104
[19]
Shende, R., Gupta, G., Patel, G. and Kumar, S. (2016) Commissioning of TrueBeamTM Medical Linear Accelerator: Quantitative and Qualitative Dosimetric Analysis and Comparison of Flattening Filter (FF) and Flattening Filter Free (FFF) Beam. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 5, 51-69. http://dx.doi.org/10.4236/ijmpcero.2016.51006
[20]
Masanga, W., Tangboonduangjit, P., Khamfongkhruea, C. and Tannanonta, C. (2016) Determination of Small Field Output Factors in 6 and 10 MV Flattening Filter Free Photon Beams Using Various Detectors. Journal of Physics Conference Series, 694, Article ID: 012027. https://doi.org/10.1088/1742-6596/694/1/012027
[21]
Feng, Z., Yue, H., Zhang, Y., Wu, H., Cheng, J. and Su, X. (2016) Monte Carlo Simulation of Beam Characteristics from Small Fields Based on TrueBeam Flattening-Filter-Free Mode. Radiation Oncology, 11, Article No. 30.
https://doi.org/10.1186/s13014-016-0601-2
[22]
Sharma, D.S., Chaudhary, R.K., Sharma, S.D., Pilakkal, S., Rasal, S.K., Sawant, M.B., et al. (2017) Experimental Determination of Stereotactic Cone Size and Detector Specific Output Correction Factor. The British Journal of Radiology, 90, Article ID: 20160918. https://doi.org/10.1259/bjr.20160918
[23]
Qin, Y.J., Zhong, H.L., Wen, N., Snyder, K., Huang, Y.M. and Chetty, I.J. (2016) Deriving Detector-Specific Correction Factors for Rectangular Small Fields Using a Scintillator Detector. Journal of Applied Clinical Medical Physics, 17, 379-391.
https://doi.org/10.1120/jacmp.v17i6.6433
[24]
Koger, B., Price, R., Wang, D., Toomeh, D., Geneser, S. and Ford, E. (2020) Impact of the MLC Leaf-Tip Model in a Commercial TPS: Dose Calculation Limitations and IROC-H Phantom Failures. Journal of Applied Clinical Medical Physics, 21, 82-88. https://doi.org/10.1002/acm2.12819
[25]
Kim, J., Han, J.S., Hsia, A.T., Li, S., Xu, Z. and Ryu, S. (2018) Relationship between Dosimetric Leaf Gap and Dose Calculation Errors for High Definition Multi-Leaf Collimators in Radiotherapy. Physics and Imaging in Radiation Oncology, 5, 31-36.
https://doi.org/10.1016/j.phro.2018.01.003
[26]
Mzenda, B., Mugabe, K.V., Sims, R., Godwin, G. and Loria, D. (2014) Modeling and Dosimetric Performance Evaluation of the RayStation Treatment Planning System. Journal of Applied Clinical Medical Physics, 15, 29-46.
https://doi.org/10.1120/jacmp.v15i5.4787