全部 标题 作者
关键词 摘要

OALib Journal期刊
ISSN: 2333-9721
费用:99美元

查看量下载量

相关文章

更多...

Correlation between GEM Premier 3000 and Vitros5.1+5600, SYSMEX XN-9000 in Detecting Electrolytes and Red Blood Cell Volume

DOI: 10.4236/jbise.2022.151005, PP. 44-50

Keywords: Blood Gas Analyzer, Serum Potassium, Serum Sodium, Blood Glucose, Hemoglobin, Red Blood Cell Volume, Central Laboratory

Full-Text   Cite this paper   Add to My Lib

Abstract:

Objective: To explore the consistency and relevance of the results of the bedside rapid blood gas analyzer GEM premier 3000, the Vitros5.1+5600 biochemical immunoassay analyzer and the SYSMEX XN-9000 automatic blood cell analyzer in the central laboratory detecting serum potassium (K+), serumsodium (Na+), blood glucose (Glu), hemoglobin (Hb) and red blood cell volume (Hct). And to provide a reference for the accurate interpretation of the bedside blood gas analysis report. Method: Usually, ICU patients will be taken arterial blood gas, biochemical and blood samples through the arterial indwelling needle; at the same time patients’ potassium, serum sodium, blood glucose, hemoglobin and red blood cell volume will be detected. This study implemented paired t-test and correlation regression analysis on each group of data, and used the analysis quality requirements (allowable total error) of CLIA’88 proficiency testing program as the criteria for clinical acceptance. Results: The paired t-test showed that the serum potassium, serum sodium and blood glucose detected by GEM premier 3000 and Vitros5.1+5600 were significantly different; and the hemoglobin and red blood cell volume detected by GEM premier 3000 and SYSMEX XN-9000 were significantly different (P < 0.05). The Pearson correlation coefficients (r) of hemoglobin, red blood cell volume and red blood cell volume were 0.860, 0.886, 0.924, 0.841 and 0.856, respectively, and the above test items all had good correlations (P < 0.05). The average (SE) of the paired differences of K+, Na+, Glu, Hb and Hct detected by the two sets of instruments is less than the allowable error of CLIA’88, and the SE of blood Na+ and Hb is less than half of the allowable error of CLIA’88. Conclusion: The test results of GEM premier 3000, the central laboratory Vitros5.1+5600 and SYSMEX XN-9000 have good correlation, but the consistency is not good. The test results of GEM premier 3000 cannot replace the central laboratory.

References

[1]  1.Luo, K., Liang, S., Liu, Y., He, H., He, Y. and Huang, H. (2019) Correlation and Consistency of Electrolytes, Blood Glucose and Lactic Acid Test Results Detected by POCT Blood Gas Analyzer and Central Laboratory. International Journal of Laboratory Medicine, 40, 1938-1940+1944.
[2]  Zatloukal, J., Pouska, J., Kletecka, J., Pradl, R. and Benes, J. (2016) Comparison of the Accuracy of Hemoglobin Point of Care Testing Using Hemocue and GEM Premier 3000 with Automated Hematology Analyzer in Emergency Room. Journal of Clinical Monitoring and Computing, 30, 949-956.
https://doi.org/10.1007/s10877-015-9799-z
[3]  Zhang, L., Zhu, Y., Ran, Y. and He, X. (2020) Discussion on the Consistency of Blood Gas Analyzer and Conventional Instruments in the Same Items. Chinese Journal of Laboratory Pathologist, 12, 77-80.
[4]  Chao, C.F. and Xu, Y. (2019) Research Progress on the Factors That Affect the Indwelling Time of Arterial Indwelling Needle and Nursing Countermeasures. Acta Academiae Medicinae Xuzhou, 39, 232-234.
[5]  Liu, X.M., Yang, M.L., Yi, J.Q. and Pan, X.X. (2015) Multifunctional Automatic Blood Cell Analyzer Sysmex XN-9000 Fluid Model Performance; Verification. China Medical Equipment, No. 4, 44-46.
[6]  Yu, M., Sun, Q. and Jiang, F. (2017) Performance Verification of VITROS 5600 Automatic Biochemical Analyzer. International Journal of Laboratory Medicine, 38, 56-60.
[7]  Westgard (2021) CLIA Requirements for Analytical Quality. Westgard.
https://www.westgard.com/clia.htm
[8]  Xu, J.H., Liu, D.D., Huang, X.Z. and Zhuang, J.H. (2015) Application of New Guideline CLSI EP9-A3 in Methodological Comparison and Deviation Assessment. National Medical Journal of China, 95, 894-897.
[9]  Yu, J. (2016) Investigation on the Practical Application Significance of Non-Traditionary Test Items by GEM Premier 3000 Blood Gas Analyzer. Laboratory Medicine, No. 1, 26-30.
[10]  Xiong, W.-F., Xiao, G.-J., Liu, Y.-T., Yang, N., Cao, D. and Chen, S. (2018) Comparative Analysis of the Results of Non-Traditional Detection of GEM Premier 3000Blood Gas Analyzer and Conventional Instruments in Central Laboratory. Medical Information, 31, 160-162.
[11]  Zhong, G.Z. (2020) Regression Analysis of the Difference between GEM3000 Blood Gas Analyzer and Biochemical Analyzer in the Determination of Blood Potassium. China Practical Medical, 15, 197-198.
[12]  Liu, H., Ma, Y., Zhang, P., Zhang, Y. and Yang, B. (2012) Comparison of Determination of Blood Glucose by Blood Gas Analyzer, Portable Blood Glucose Meter and Automated Biochemical Analyzer. Laboratory Medicine and Clinic, 9, 407-410.
[13]  Spielmann, N., Mauch, J., Madjdpour, C., Schmugge, M., Weiss, M. and Haas, T. (2012) Accuracy and Precision of Hemoglobin Point-of-Care Testing during Major Pediatric Surgery. International Journal of Laboratory Hematology, 34, 86-90.
https://doi.org/10.1111/j.1751-553x.2011.01363.x
[14]  Cui, F.Y. and Chen, H.W. (2017) Application Analysis of Medical Logistics System in Modern Hospital. China Medical Devices, 32, 137-140.
[15]  Li, T.M., Liang, S.X., Wang, S.X., et al. (2014) Research on Factors of Detection Accuracy of GEM Premier 3000 Blood Gas Analyzer. Hebei Medical Journal, No. 9, 1368-1370.

Full-Text

comments powered by Disqus

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133

WeChat 1538708413