Use of High Definition Multileaf Colimator for the Treatment of Trigeminal Neuralgia

DOI: 10.4236/ijmpcero.2022.113013, PP. 150-159

Keywords: Trigeminal Neuralgia, Radiosurgery, Brainstem, Brainstem-Optimized Technique (BO), Virtual Cone (VC)

Full-Text   Cite this paper   Add to My Lib

Abstract:

In the present work, a treatment technique for trigeminal neuralgia (TN) using LINAC radiosurgery is shown. The technique is based on the optimization of ten static arcs in such a way as to minimize the overlapping of the treatment fields with the brainstem. We will call this technique brainstem-optimized (BO). The results are compared with another technique described in the literature known as a virtual cone (VC). The comparison of dosimetry results that have been carried out essentially shows that the doses in the brainstem V12Gy-brainstem, D0.5cm3-brainstem and D0.035 cm3-brainstem are lower in the BO versus VC technique, and with the parameters V50% (whole brain) and V12Gy-cerebrum higher in BO versus VC. Our goal is to keep the dose to the brainstem as low as possible and, if possible, at most between 12 Gy and 15 Gy. The BO technique meets our purposes and is considered clinically acceptable at our institution.

References

[1]  Bennetto, L., Patel, N.K. and Fuller, G. (2007) Trigeminal Neuralgia and Its Management. British Medical Journal, 334, 201-205.
https://doi.org/10.1136/bmj.39085.614792.BE
[2]  Debono, B., et al. (2019) Dedicated Linear Accelerator Radiosurgery for Classic Trigeminal Neuralgia: A Single-Center Experience with Long-Term Follow-Up. World Neurosurgery, 121, e775-e785.
https://doi.org/10.1016/j.wneu.2018.09.222
[3]  Popple, R.A., et al. (2018) The Virtual Cone: A Novel Technique to Generate Spherical Dose Distributions Using a Multileaf Collimator and Standardized Control-Point Sequence for Small Target Radiation Surgery. Advances in Radiation Oncology, 3, 421-430.
https://doi.org/10.1016/j.adro.2018.02.011
[4]  Brown, T.A.D., Ayers, R.G. and Popple, R.A. (2022) Commissioning a Multileaf Collimator Virtual Cone for the Stereotactic Radiosurgery of Trigeminal Neuralgia. Journal of Applied Clinical Medical Physics, 23, e13562.
https://doi.org/10.1002/acm2.13562
[5]  Sudahar, H., Kurup, P.G.G., Murali, V. and Velmurugan, J. (2012) Dosimetric Analysis of Trigeminal Nerve, Brain Stem Doses in CyberKnife Radiosurgery of Trigeminal Neuralgia. Journal of Medical Physics, 37, 124-128.
https://doi.org/10.4103/0971-6203.99225
[6]  Klassen, P.D., Ivaturi, S.K. and Hendricks, B.K. (2022) Trigeminal Rhizotomy Using Gyroscopic Radiosurgery: A Case Report. Cureus, 14, e24951.
https://doi.org/10.7759/cureus.24951
[7]  Kano, H., et al. (2017) Estimating the Risks of Adverse Radiation Effects after Gamma Knife Radiosurgery for Arteriovenous Malformations. Stroke, 48, 84-90.
https://doi.org/10.1161/STROKEAHA.116.014825
[8]  Bentzen, S.M., et al. (2010) Quantitative Analyses of Normal Tissue Effects in the Clinic (QUANTEC): An Introduction to the Scientific Issues. International Journal of Radiation Oncology, Biology, Physics, 76, S3-S9.
https://doi.org/10.1016/j.ijrobp.2009.09.040
[9]  Tuleasca, C., et al. (2019) Stereotactic Radiosurgery for Trigeminal Neuralgia: A Systematic Review: International Stereotactic Radiosurgery Society Practice Guidelines. Journal of Neurosurgery, 130, 733-757.
https://doi.org/10.3171/2017.9.JNS17545
[10]  Wilson, T.A., Karlsson, B., Huang, L., Ramanathan, D., Oyoyo, U. and Boling, W. (2020) Optimizing Radiosurgery for Trigeminal Neuralgia: Impact of Radiation Dose and Anatomic Target on Patient Outcomes. World Neurosurg, 143, e482-e491.
https://doi.org/10.1093/neuros/nyaa447_556
[11]  Lettmaier, S. (2014) Radiosurgery in Trigeminal Neuralgia. Physica Medica: European Journal of Medical Physics, 30, 592-595.
https://doi.org/10.1016/j.ejmp.2014.05.006
[12]  Timmerman, R.D. (2008) An Overview of Hypofractionation and Introduction to This Issue of Seminars in Radiation Oncology. Seminars in Radiation Oncology, 18, 215-222.
https://doi.org/10.1016/j.semradonc.2008.04.001
[13]  Rose, M.S., et al. (2020) Multi-Institution Validation of a New High Spatial Resolution Diode Array for SRS and SBRT Plan Pretreatment Quality Assurance. Medical Physics, 47, 3153-3164.
https://doi.org/10.1002/mp.14153
[14]  Xu, Q., et al. (2022) Implementing and Evaluating a High-Resolution Diode Array for Patient-Specific Quality Assurance of Robotic Brain Stereotactic Radiosurgery/Ra-diotherapy. Journal of Applied Clinical Medical Physics, 23, e13569.
https://doi.org/10.1002/acm2.13569

Full-Text

Contact Us

[email protected]

QQ:3279437679

WhatsApp +8615387084133