Single
isocenter multiple target stereotactic radiosurgery (SIMT-SRS) has potentially
emerged as a new pillar in radio-immune combination therapy for the management
of brain metastasis. Accuracy and efficiency are pushed to a higher level in the era of the linear accelerator-based SIMT-SRS. This short review focuseson patient selection, image preparation, patient
simulation, electronic portal imaging device (EPID)
QA, and the patient treatment process in the SIMT-SRS treatment only. Image-relevant recommendations and guidelines are presented
and contrast application, acquisition efficiency, and alignment accuracy of CT
and MRI images are explored. With guidance, the SIMT-SRS can be implemented with high precision and efficiency. 1mm or
0.5 mm and non-uniform PTV margin expansion for all targets would
become possible. It will enhance cancer killing effect in radio-immune
combination therapy. General routine daily, monthly, and annual linear
accelerator image quality assurancesare excluded.
References
[1]
Liu, Q., Tong, X.Z. and Wang, J.F. (2019) Management of Brain Metastases: History and the Present. Chinese Neurosurgical Journal, 5, Article No. 1. https://doi.org/10.1186/s41016-018-0149-0
[2]
Ngwa, W., Irabor, Q.C., Schoenfeld, J.D., Hesser, J., Demaria, S. and Formenti, S.C. (2018) Using Immunotherapy to Boost the Abscopal Effect. Nature Reviews Cancer, 18, 313-322. https://doi.org/10.1038/nrc.2018.6
[3]
Klein, E.E., Hanley, J., Bayouth, J., Yin, F.F., et al. (2009) Task Group 142 Report: Quality Assurance of Medical Accelerators. Medical Physics, 36, 4197-4212. https://doi.org/10.1118/1.3190392
[4]
Hanley, J., Dresser, S., Simon, W., et al. (2021) AAPM Task Group 198 Report: An Implementation Guide for TG 142 Quality Assurance of Medical Accelerators. Medical Physics, 48, e830-e885. https://doi.org/10.1002/mp.14992
[5]
Becker, S.J., Lipson, E.J., Jozsef, G., Molitoris, J.K., et al. (2023) How Many Brain Metastases Can Be Treated with Stereotactic Radiosurgery before the Radiation Dose Delivered to Normal Brain Tissue Rivals That Associated with Standard Whole Brain Radiotherapy? Journal of Applied Clinical Medical Physics, 24, e13856. https://doi.org/10.1002/acm2.13856
[6]
Price, R., Allison, J. and Clarke, G. (2015) 2015 Magnetic Resonance Imaging Quality Control Manual. American College of Radiology.
[7]
Lu, L.C., Yang, X.Y., Raterman, B., et al. (2023) Assessment of MRI Image Distortion Based on 6 Consecutive Years of Annual QAs and Measurements on 14 MRI Scanners Used for Radiation Therapy. Journal of Applied Clinical Medical Physics, 24, e13843. https://doi.org/10.1002/acm2.13843
[8]
Xiao, Y.D., Paudel, R., Liu, J., Ma, C., Zhang, Z.S. and Zhou, S.K. (2016) MRI Contrast Agents: Classification and Application (Review). International Journal of Molecular Medicine, 38, 1319-1326. https://doi.org/10.3892/ijmm.2016.2744
[9]
Kaufmann, T.J., Smits, M., Boxerman, J., et al. (2020) Consensus Recommendations for a Standardized Brain Tumor Imaging Protocol for Clinical Trials in Brain Metastases. Neuro-Oncology, 22, 757-772. https://doi.org/10.1093/neuonc/noaa030
[10]
Derks, S.H.A.E., Veldt, A.A.M. and Smits, M. (2022) Brain Metastases: The Role of Clinical Imaging. British Journal of Radiology, 95, Article 20210944. https://doi.org/10.1259/bjr.20210944
[11]
Huq, M.S., Fraass, B.A., Dunscombe, P.B., et al. (2016) The Report of Task Group 100 of the AAPM: Application of Risk Analysis Methods to Radiation Therapy Quality Management. Medical Physics, 43, 4209-4262. https://doi.org/10.1118/1.4947547
[12]
Samei, E., Bakalyar, D., Boedeker, K.L., et al. (2019) Performance Evaluation of Computed Tomography Systems: Summary of AAPM Task Group 233. Medical Physics, 46, e735-e756. https://doi.org/10.1002/mp.13763
[13]
Bae, K.T., Heiken, J.P. and Brink, J.A. (1998) Aortic and Hepatic Contrast Medium Enhancement at CT Effect of Reduced Cardiac Output in a Porcine Model. Radiology, 207, 657-662. https://doi.org/10.1148/radiology.207.3.9609887
[14]
Bae, K.T. (2010) Intravenous Contrast Medium Administration and Scan Timing at CT: Considerations and Approaches. Radiology, 256, 32-61. https://doi.org/10.1148/radiol.10090908
[15]
Gao, J.F. and Liu, X.Q. (2016) Off-Isocenter Winston-Lutz Test for Stereotactic Radiosurgery/Stereotactic Body Radiotherapy. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 5, 154-161. https://doi.org/10.4236/ijmpcero.2016.52017
[16]
Gao, J.F. and Liu, X.Q. (2019) Winston-Lutz-Gao Test on the True Beam STx Linear Accelerator. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 8, 9-20. https://doi.org/10.4236/ijmpcero.2019.81002
[17]
Dogan, N., Mijnheer, B.J., Padgett, K., et al. (2023) AAPM Task Group Report 307: Use of EPIDs for Patient-Specific IMRT and VMAT QA. Medical Physics, 50, e865-e903. https://doi.org/10.1002/mp.16536
[18]
Brocka, K.K., Mutic, S., McNutt, T.R., Li, H. and Kessler, M.L. (2017) Use of Image Registration and Fusion Algorithms and Techniques in Radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No. 132. Medical Physics, 44, e43-e76. https://doi.org/10.1002/mp.12256