全部 标题 作者
关键词 摘要

OALib Journal期刊
ISSN: 2333-9721
费用:99美元

查看量下载量

相关文章

更多...

Determination of biphenyl and biphenyl oxide in aqueous samples by headspace single drop microextraction coupled to gas chromatography

DOI: 10.1590/S0103-50532012000400004

Keywords: biphenyl, biphenyl oxide, single drop microextraction, headspace.

Full-Text   Cite this paper   Add to My Lib

Abstract:

in this work, biphenyl and biphenyl oxide were extracted by headspace single drop microextraction (hs-sdme) and analyzed by gas chromatography-flame ionization detection (gc-fid). the extraction occurred from the tip of a microsyringe by suspending a 3.5 μl drop of toluene, as extracting solvent, containing acetonaphton, as internal standard. the effect of different parameters such as the nature of extraction solvent, microdrop and sample temperatures, stirring rate, microdrop and sample volumes, ionic strength and extracting time on the extraction efficiency of the analytes were investigated and optimized. under optimized conditions the limits of detection (s/n = 3) of biphenyl and biphenyl oxide were 0.40 ± 0.03 and 0.30 ± 0.06 μg ml?1, respectively. good linearities were obtained for both analytes with the correlation coefficients higher than 0.997 and the relative standard deviations (r.s.d.) were in the range of 1-3%. the recoveries of analytes from spiked water samples were near to 100%.

Full-Text

Contact Us

[email protected]

QQ:3279437679

WhatsApp +8615387084133