Materials  2014 

Gain of Imaging Fidelity by Employing a Higher Number of Independent Transmit Channels Together with Slice-Selective Radio-Frequency (RF) Shimming at 7T

DOI: 10.3390/ma7010030

Keywords: image homogeneity, radio frequency shimming, RF shimming, RF-power control, transmit array, multi-slice, 7T

Full-Text   Cite this paper   Add to My Lib

Abstract:

Dielectric resonance effects and radio-frequency (RF) power deposition have become challenging issues for magnetic resonance imaging at ultrahigh-field (UHF) strengths. The use of transmit (Tx) coil arrays with independently-driven RF sources using a parallel transmission system is a promising method for alleviating the resulting RF inhomogeneities. In this study, the effect on homogeneity and RF-power when employing a higher number of transmit channels with multi-slice acquisition in vivo at high field strength (7T) is scrutinized. An 8-channel head coil array was driven to emulate circular polarized (CP) and 2-, 4-, and 8-channel independent transmit configurations at 7T. Static RF shimming was employed on human subjects in order to homogenize the B 1 + field in the excited volume. Slice-selective and global RF shimming methods were applied with CP and 2-, 4-, and 8-channel transmit channel configurations. RF shimming was performed from CP to 2-, 4-, and 8-channel Tx configurations globally and slice-selectively. Systematic improvement in B 1 + homogeneity and/or reduction in RF-power were observed. RF shimming in the human brain with 8-channel transmit and slice-selective shimming yields an increase in B 1 + homogeneity of 43% and/or reduces RF-power by 68% when compared with CP global RF shimming at 7T.

References

[1]  Van de Moortele, P.-F.; Akgun, C.; Adriany, G.; Moeller, S.; Ritter, J.; Collins, C.M.; Smith, M.B.; Vaughan, J.T.; U?urbil, K. B1 destructive interferences and spatial phase patterns at 7T with a head transceiver array coil. Magn. Reson. Med?2005, 54, 1503–1518.
[2]  Vaughan, T.; Ibrahim, T.S.; Hue, Y.-K.; Tang, L. Understanding and manipulating the RF fields at high field MRI. NMR Biomed?2009, 22, 927–936.
[3]  Levitt, M.H.; Freeman, R. NMR population inversion using a composite pulse. J. Magn. Reson?1979, 33, 473–476.
[4]  Tannus, A.; Garwood, M. Adiabatic pulses. NMR Biomed?1997, 10, 423–434.
[5]  Adriany, G.; van de Moortele, P.-F.; Wiesinger, F.; Moeller, S.; Strupp, J.P.; Andersen, P.; Snyder, C.; Zhang, X.; Chen, W.; Pruessmann, K.P.; et al. Transmit and receive transmission line arrays for 7 Tesla parallel imaging. Magn. Reson. Med?2005, 53, 434–445.
[6]  Metzger, G.J.; Snyder, C.; Akgun, C.; Vaughan, T.; Ugurbil, K.; van de Moortele, P.-F. Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject-dependent transmit phase measurements. Magn. Reson. Med?2008, 59, 396–409.
[7]  Collins, C.M.; Liu, W.; Swift, B.J.; Smith, M.B. Combination of optimized transmit arrays and some receive array reconstruction methods can yield homogeneous images at very high frequencies. Magn. Reson. Med?2005, 54, 1327–1332.
[8]  Ibrahim, T.; Lee, R.; Baertlein, B.; Abduljalil, A.; Zhu, H.; Robitaille, P. Effect of RF coil excitation on field inhomogeneity at ultra high fields: A field optimized TEM resonator. Magn. Reson. Imaging?2001, 19, 1339–1347.
[9]  Mao, W.; Smith, M.B.; Collins, C.M. Exploring the limits of RF shimming for high-field MRI of the human head. Magn. Reson. Med?2006, 56, 918–922.
[10]  Willinek, W.A.; Gieseke, J.; Kukuk, G.M.; Nelles, M.; Konig, R.; Morakkabati-Spitz, N.; Traber, F.; Thomas, D.; Kuhl, C.K.; Schild, H.H. Dual-source parallel radiofrequency excitation body MR imaging compared with standard MR imaging at 3.0 T: Initial clinical experience. Radiology?2010, 256, 966–975.
[11]  Adriany, G.; Auerbach, E.J.; Snyder, C.J.; G?zübüyük, A.; Moeller, S.; Ritter, J.; van de Moortele, P.-F.; Vaughan, T.; U?urbil, K. A 32-channel lattice transmission line array for parallel transmit and receive MRI at 7 Tesla. Magn. Reson. Med?2010, 63, 1478–1485.
[12]  Setsompop, K.; Alagappan, V.; Gagoski, B.; Witzel, T.; Polimeni, J.; Potthast, A.; Hebrank, F.; Fontius, U.; Schmitt, F.; Wald, L.L.; et al. Slice-selective RF pulses for in vivo B1+ inhomogeneity mitigation at 7 Tesla using parallel RF excitation with a 16-element coil. Magn. Reson. Med?2008, 60, 1422–1432.
[13]  Vernickel, P.; R?schmann, P.; Findeklee, C.; Lüdeke, K.-M.; Leussler, C.; Overweg, J.; Katscher, U.; Gr?sslin, I.; Schünemann, K. Eight-channel transmit/receive body MRI coil at 3T. Magn. Reson. Med?2007, 58, 381–389.
[14]  Snyder, C.J.; DelaBarre, L.; Moeller, S.; Tian, J.; Akgun, C.; van de Moortele, P.-F.; Bolan, P.J.; Ugurbil, K.; Vaughan, J.T.; Metzger, G.J. Comparison between eight- and sixteen-channel TEM transceive arrays for body imaging at 7T. Magn. Reson. Med?2012, 67, 954–964.
[15]  Van de Moortele, P.-F.; Akgun, C.; Snyder, C.; DelaBarre, L.; Adriany, G.; Vaughan, T.; Ugurbil, K. Impact of Coil Diameter and Number of Coil Elements on B1 Destructive Interferences with Stripline Coil Arrays at 7 Tesla, Presented at ISMRM 15th Annual Meeting, Seattle, Washington, USA, 6–12 May 2006.
[16]  Wu, X.; Sebasian, S.; Adriany, G.; Auerbach, E.J.; Ugurbil, K.; van de Moortele, P.-F. Enhanced Whole Brain Excitation Perfomance of Parallel Transmission with a Z-Encoding RF Coil Array at 7T, Presented at ISMRM 20th Annual Meeting, Melbourne, Australia, 5–11 May 2012.
[17]  Childs, A.S.; Malik, S.J.; O’Regan, D.P.; Hajnal, J.V. Impact of number of channels on RF shimming at 3T. Magn. Reson. Mater. Phys?2013, 26, 401–410.
[18]  Darji, N.; Kim, K.-N.; Patel, G.; Fautz, H.-P.; Bernarding, J.; Speck, O. Presented at ISMRM 19th Annual Meeting, Montréal, QC, Canada, 10 May 2011.
[19]  Curtis, A.T.; Gilbert, K.M.; Klassen, L.M.; Gati, J.S.; Menon, R.S. Slice-by-slice B1+ shimming at 7T. Magn. Reson. Med?2012, 68, 1109–1116.
[20]  Setsompop, K.; Wald, L.; Alagappan, V.; Gagoski, B.; Adalsteinsson, E. Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 Tesla with eight channels. Magn. Reson. Med?2008, 59, 908–915.
[21]  Schneider, J.T.; Kalayciyan, R.; Haas, M.; Herrmann, S.R.; Ruhm, W.; Hennig, J.; Ullmann, P. Inner-volume imaging in vivo using three-dimensional parallel spatially selective excitation. Magn. Reson. Med?2013, 69, 1367–1378.
[22]  Gilbert, K.M.; Belliveau, J.-G.; Curtis, A.T.; Gati, J.S.; Klassen, L.M.; Menon, R.S. A conformal transceive array for 7T neuroimaging. Magn. Reson. Med?2012, 67, 1487–1496.
[23]  Collins, C.M.; Wang, Z.; Smith, M.B. A Conservative Method for Ensuring Safety within Transmit Arrays, Presented at ISMRM 16th Annual Meeting, Berlin, Germany, 19–25 May 2007.
[24]  Fautz, H.-P.; Vogel, M.; Gross, P.; Kerr, A.; Zhu, Y. B1 Mapping of Coil Arrays for Parallel Transmission, Presented at ISMRM 17th Annual Meeting, Toronto, ON, Canada, 3–9 May 2008.
[25]  Klose, U. Mapping of the radio frequency magnetic field with a MR snapshot FLASH technique. Med. Phys?1992, 19, 1099–1104.
[26]  Pinkerton, R.G.; McKinnon, G.C.; Menon, R.S. SENSE optimization of a transceive surface coil array for MRI at 4T. Magn. Reson. Med?2006, 56, 630–636.